Putting into use
Open valves carefully and chek installation for
leakage.
If the sleep mode of the counter is enabled
(Display: SLEEP 1), then it must be deacti-
vated by longer pressing the button (>5s).
While the system is operating, check whether
the volume display advances and the temper-
atures displayed correspond with the actual
temperatures (see the display overview).
Wait for the temperature display to be updated
(1-2 sec).
Secure meter with the enclosed sealing mate-
rial against unauthorised removal.
Fill in the putting into use report in accordance
with PTB-Directive TR K9.
Mounting in existing immersion sleeves:
The device C5 can be put into use in connection
with existing immersion sleeves in accordance
with the article „Putting into use of MID homolo-
gated temperature sensors” released in the PTB
notications 119 (2009), vol. 6.
Based on current information, the regulation
has a period of validity until 30.10.2016. For the
identication and marking of the usable existing
immersion sleeves in connection with the C5
device, an identication and marking set can be
delivered from our company.
Consumer
Symmetrical sensor installation for zelsius
®
C5-IUF with the return
sensor integrated in the measuring capsule
Supply
Return
Consumer
Symmetrical sensor installation for zelsius
®
C5-IUF
Supply
Return
Consumer
Symmetrical sensor installation for zelsius
®
C5-IUF with the return
sensor integrated in the measuring capsule
Supply
Return
Consumer
Symmetrical sensor installation for zelsius
®
C5-IUF
Supply
Return